Problems we were actually asked to solve

Customer names are withheld under NDA. Dimensions, tolerances and outcomes are reported as delivered.

Painted and engraved cast-acrylic mask plate for a laser system
Full case study

Twenty H7 bores at Ø0.02 mm true position — on a painted, engraved acrylic plate

The complete 14-operation route for a laser-systems OEM: thermal budget, stress control, optical windows, masked painting and engraving — plus the six drawing questions closed before the first cut, and the measured results on every requirement.

Read the full case study →
Five-layer bonded PMMA microfluidic chip Microfluidics · IVD

5-layer PMMA chip holds ±8 µm channel depth at series scale

Challenge. A point-of-care developer had a working prototype machined in-house, but channel depth drifted between lots and thermal bonding occasionally collapsed the shallower channels.

What we changed. We switched to stress-relieved cast PMMA, adjusted the bonding profile to a two-stage ramp with controlled pressure, and added in-process depth measurement on 100 % of wafers before bonding.

Result. Depth held to ±8 µm over a 75 × 25 mm footprint, burst pressure above 4 bar, and a first production lot shipped in 21 days.

±8 µmChannel depth
4 barBurst pressure
21 daysFirst lot
Diamond fly-cut acrylic light guide Optics · Diagnostics

Acrylic light guide replaces glass at one third of the cost

Challenge. A reader manufacturer was using a ground glass light guide. Cost and a fragile supply chain pushed them to look for a polymer alternative, but early acrylic attempts lost 15 % transmission at the coupling face.

What we changed. Diamond fly-cutting of both end faces to Ra 0.02 µm, edge blackening to control stray light, and a fixture that keeps the part flat during polishing.

Result. 92 % transmission across the working band, part cost reduced by 65 %, no change to the optical model in their firmware.

Ra 0.02 µmSurface finish
92 %Transmission
−65 %Part cost
Bonded acrylic flow cell with quartz windows Water · Analytical

In-line flow cell survives 10,000 pressure cycles

Challenge. A turbidity sensor maker was losing units in the field to window debonding. The acrylic body bonded to quartz windows failed at the adhesive interface under thermal cycling.

What we changed. Moved to a two-component adhesive with matched CTE, added a mechanical register so the bond line carries no shear, and introduced 100 % pressure-decay testing.

Result. Zero leak failures across 10,000 accelerated cycles at 6 bar; field return rate dropped from 3.1 % to under 0.2 %.

10 kPressure cycles
0Leak failures
<0.2 %Field returns
Multi-port acrylic manifold under inspection Valves · Manifolds

24-port manifold machined to ±15 µm true position

Challenge. A fluidics integrator needed 24 ports on a single acrylic block, aligned well enough for push-to-connect fittings to seal without thread sealant.

What we changed. Single-setup machining on a granite-bed mill with a custom fixture, drilling sequenced to avoid thermal drift, and CMM verification of every port position on the CMM before assembly.

Result. True position within ±15 µm across the face, assembly leak test pass rate 99.6 %, and a repeatable 12-day cycle for series orders.

±15 µmTrue position
99.6 %Leak test pass
12 daysSeries cycle